ST-LAB7 Automatic Torque-Force-Pressure Tester
Sure Torque’s ST-LAB7 is multi-purpose full automatic force-torque-pressure tester with FDA CFR21 part 11 compliant user management and audit logging. The instrument is optimized for performing test on:
– Syringes and containers with Luer Lock – ISO 80369, ISO 11040 including pressure tests
– Injector pen devices (Needle-based injection systems for medical use – ISO 11608)
Additional Information
Included accessories |
1 set of change parts (container clamp and cap collet) Pc software and communication cables Calibration kit Foamed heavy duty transport case |
Smartphone requirement | No |
Android version | No |
Measuring range | +/- 3 Nm / 250 N / 500 kPa |
Units of measurement | LbfIn, LbfFt, Nm, dNm, Ncm, kgfcm / N, Kg, Lb / kPa |
Displayed values | Actual, Max1, (Max2), torque-time graph |
Accuracy | +/- 0,2% F.S. |
Safe Overload | 120% |
Maximum measuring frequency | 350 Hz (2.85 ms) |
Range analysis | Yes – In range / under range / over range |
Capable closure types | any kind of twist cap |
Maximum sample size | D100 x 300 mm |
Calibration kit | included |
Operation time | No |
DC supply | DC 24V with included AD110/230 V adapter |
Bluetooth range | No |
Included software | Data manager for PC |
Memory | Unlimited – all te results of the life-cycle are stored |
Stored data’s | Batch & measurement nr, peak results, date, time, graph, instrument serial number, comment, calibration date, BATCH code, IPC code |
Resistance grade | IP 54 |
Language | English, German, Hungarian + custom language |
Dimensions: Footprint / Height | 200x360x480 mm |
Net weight | 25 kg |
Shipping packaging | Heavy duty waterproof case as option |
Shipping dimensions | 35x37x90 cm |
Shipping weight | depends on packing |
Country of Manufacture | Hungary |
Warranty | 3 years |
MEASURING WITH THE ST-LAB
The ST-LAB is a user-friendly one-button operation unit, controlled by a PLC.
• Choose a sample type on the touch screen.
• Choose the desired test method on the PLC by using the PLC’s navigation buttons.
• Place the sample (and close the safety door if there is)
• Hit the activation buttons to start a test cycle.
o The container clamp closes and holds the container tight during the test cycle
o The testing head sinks to the cap level and grabs the cap
o The desired test method will be completed
o The sample will be released
All manipulation forces can be set on a wide range so that you can choose the forces matching the best to your samples and test requirements.
• You can start a new test by hitting the activation buttons
The speed of the test cycle can be set on a wide range, however the slower a test cycle is, the more accurate it is. The internal memory of the plc stores all results of the machine life time with graph and all details which can be imported into a pc. Results can be monitored by on line data collector too.
Basic test options:
1. Annex F (normative) Resistance to separation from axial load test method
2. Annex G (normative) Resistance to separation from unscrewing test method
3. Annex H (normative) Resistance to overriding test method
4. Annex I (normative) Disconnection by unscrewing test method
5. Cap release torque test
6. Tensile test
Optional modes (pressure test option)::
7. Annex B (normative) *Leakage by pressure decay test method
8. Annex C (normative) Falling drop positive-pressure liquid leakage test method
9. Annex D (normative) Subatmospheric-pressure air leakage test method